BSA (Bovine Serum Albumin) Assay Kit

GoodsCode: BF03097-96T
Save
Specs Price Futures Quantity
  • 96T
    390.00
    Ask

Product Informations

  • Product No. BF03097-96T
  • Product Name BSA (Bovine Serum Albumin) Assay Kit
  • Category 446
  • Instruction Manual /files/uploads/instructions/2024/202409/BF03097_BSA(牛血清白蛋白)检测试剂盒_说明书.pdf

Picture

Product Details

1. Introduction to the Kit
This quantitative detection kit for bovine serum albumin (BSA) enzyme-linked immunosorbent assay (ELISA) uses monoclonal paired antibodies against BSA, which are respectively used as coating antibodies and detection antibodies to form a double antibody sandwich ELISA detection kit. It is used as a reference standard with the BSA standard of the American Bureau of Standards. This kit can detect a minimum of 1ng/mL of BSA.
2. Kit Components:

 
Name Quantity (specification Storage temperature
96-well slats coated with anti-BSA monoclonal antibody 1 bottle 4℃
BSA reference standards: 32ng/mL, 16ng/mL, 8ng/mL, 4ng/mL, 2ng/mL, 1ng/mL, 0ng/mL  1 bottle(0.4mL/bottle) -20/-80
Internal reference of BSA (20ng/mL) : 1 bottle(0.4mL/瓶)  
TMB substrate chromogenic solution 1 set(A 和 B 各 6.5mL) 4℃
Chromogenic substrate stop solution (2N, H2SO4) : 1 bottle(13mL) 4℃
Sample diluent 1 bottle(13mL) 4℃
Washing solution (10 times concentrated) : 1 bottle(50mL) 4℃
Instruction Manual 1 copy 4℃
Enzyme-labeled anti-BSA antibody 1 bottle(0.13mL) -20/-80
Enzyme-labeled antibody diluent 1 bottle(13mL) 4℃
3. Required instruments:
Microplate reader (wavelength 450nm, 10μL, 100μL, 1mL pipettes, 300μl multi-channel pipettes, pipette tips, 37℃ constant temperature water bath.)
4. Operating Steps:
4.1 Dilute the concentrated washing solution 10 times with distilled water.
4.2 Take out the 96-well microplate coated with anti-BSA and place it at room temperature to warm up. Prepare two parallel Wells for each concentration of the standard substance and add 50μL per well to the microplate. Add the sample to be tested and the internal reference, two Wells each, 50μL per well. Then, dilute the enzyme-labeled antibody with the enzyme-labeled antibody dilution solution at a ratio of 1:100 and add it to each reaction well, with 100μL per well (dilute as much as needed, prepare as needed). The sample addition process must be completed within 15 minutes.
4.3 Place at room temperature of 20 to 24 degrees Celsius and react in the dark for 1 hour. Wash four times with the washing solution and pat dry.
4.4 Add 100μL of TMB substrate chromogenic solution per well with a pipette. Do not shake. React at room temperature of 20 to 24 ° C in the dark for 10 minutes.
4.5 Add 100μL of stop solution per well with a pipette and shake slightly for 5 seconds to mix well.
4.6 Read the value at a wavelength of 450nm on the microplate reader. 5. Precautions
5.1 If the BSA content in the sample is higher than 32ng/mL, the sample can be appropriately diluted.
5.2 If the sample is a freeze-dried preparation, it should be thoroughly dissolved and mixed with the sample diluent.
5.3 During the reaction, the entire plate should be placed in a clean and sealed space to prevent foreign objects from falling into the reaction holes.
After washing and drying, any residual air bubbles in each hole should be eliminated.
After opening, put the unused strips back into the sealed plastic bag with desiccant and store it in a sealed container at 4℃.

5.6 During the operation process, the pipette tips, washers (bottles), vessels, etc. used should all be strictly protected from contamination by environments and items with BSA.

6. Result Determination
6.1 Valid experimental parameters:
The experimental results are valid only if the following valid parameters are met:
The OD value of 0ng/ml is ≤0.1. The OD value of 32ng/ml is ≥0.80. Positive reference: 18-22ng/ml; R2≥0.98;
6.2 Calculation:

In EXCEL, a scatter plot is made with the standard concentration as the abscissa and the zeroing-out OD value as the ordinate. A trend line is added, and then the curve equation and R2 value are displayed. After zeroing the OD value of the test sample, it is substituted into the linear equation to calculate the corresponding concentration.
 
Standard concentration OD1 OD2 OD3 Average value Zero Elimination OD
0 0.069 0.062 0.064 0.065 0
1 0.111 0.109 0.113 0.111 0.046
2 0.13 0.127 0.131 0.129 0.064
4 0.191 0.192 0.193 0.192 0.127
8 0.332 0.319 0.327 0.326 0.261
16 0.583 0.588 0.596 0.589 0.524
32 1.039 1.025 1.054 1.039 0.974
 
6.4 Judgment:

In the standard curve, let OD0 to 16ng/mL be the interval A; OD16 to 32ng/mL is interval B; Interval C is greater than OD32 ng/mL.

6.1.1 If the measured OD value is less than the OD value corresponding to 2ng/mL in the standard curve, the determination result of BSA content shall be counted as less than 2ng/mL.
6.1.2 If the OD values measured for samples of different dilution concentrations all fall within interval A, the final result shall be calculated based on the measured value of the lowest dilution.
6.1.3 If the measured values of different dilutions are all within interval B or both within interval A and B or both within interval B and C, the measured value closest to OD16ng/mL shall be taken.
6.1.4 If the measured values at different dilutions are all within the range C, the sample needs to be further diluted for measurement to make its measured value fall within the OD value range corresponding to 2 to 32ng/mL. The calculation method shall be as described in 6.1.2-4 above.
6.1.5 If the OD of the sample measured at high dilution is higher than that of the sample at low dilution or undiluted, it may be due to operational error or excessive BSA content in the sample (HOOK effect), and the dilution should be retried or adjusted.
6.1.6 The final result should take into account the dilution factor (actual BSA content = calculated value X dilution factor).

7. Technical Specifications of this kit
7.1 Detection limit and quantification limit
Detection limit: Zero concentration standard solution plus twice the standard deviation (calculated values of 20 zero Wells). The detection limit of this kit is calculated as 0.2ng/mL(ppb). The quantification limit is the lower detection limit of the standard curve, 1 ng/mL.
7.2 Accuracy and precision
Calculate the intra-batch and inter-batch coefficients of variation. Five repeated tests were conducted at high, medium and low concentrations (20, 10, 5) to calculate the intra-batch coefficients of variation. The inter-batch coefficients of variation were calculated three times at different times, and the CV% was all less than 10%.
7.3 Cross-reaction

Concentration(ng/ml) SD recovery rate% CV% Concentration(ng/ml) SD recovery rate% CV%
20 0.06 107.3 5.6 20 0.047 102.6 4.6
10 0.049 105.4 4.7 10 0.022 101.2 2.2
5 0.059 99.66 5.9 5 0.028 97.6 2.9
This kit has no cross-reaction with human albumin, mouse serum, rabbit serum and fish gelatin, but has a 10% cross-reaction with goat serum
(The percentage of the ratio of BSA to the concentration of goat serum at the same OD value shall be taken as the standard).
Substances that do not cross-react with BSA
Reaction substance Reaction concentration
Human blood albumin 10mg/ml
Mouse serum 10%
Rabbit serum 10%
Fish gelatin 1%

7.4 Hook Effect
When the sample concentration is much higher than the highest concentration on the standard curve, a HOOK effect will occur, that is, when the reaction OD value of BSA is less than 32ng/mL, the concentration is above 500μg/mL for this kit.

Citation (0)

Q1: BSA detection kit

Related Products

内容已折叠

Recently Viewed Products

If you are interested in our products or services, you can contact us through the following methods

  • In the form below Describe Your Needs
  • Click "Online Consultation" on the right side of the page for quick assistance Quick Tool Consultation
  • Call our toll-free number 400-921-8278
  • Send an email to our mailbox sales@biodragon.cn
« »